Circulation Type
Natural Convection
Operating temperature range
Room Temp. +5 to 260°C
Temp. adjustment accuracy
±1.0°C (at 260°C )
Temp. fluctuation
3.0°C (at 260°C )
Temp. distribution accuracy
±5.0°C (at 260°C )
Temp. gradient
20°C (at 260°C)
Max. Temp. Reaching Time
Approx. 70 min
Interior Material
Stainless Steel
Exterior Material
Cold Rolled Steel Plate with Melamine Resin Baked Finish
Heat Insulating Material
Glass Wool
Heater
SUS Pipe heater 1.2kW
Sensor
K-Thermocouple (for temp. adjustment and overheat prevention)
Cable Hole
I.D. 33mm (Right Side)
Exhaust Port
I.D. 33mm x 2 (Top Side)
Observation Window
250 x 280mm Tempered Glass x 2 pcs
Temp. Control System
PID Control by Micro Computer
Temp. Setting
Digital Setting by UP/DOWN Key
Temp. Display
Digital Display by Green LED
Other Display
Temp. Pattern Display for Operating Display by LED
Timer
1min. to 9hrs.59min.
Operating Function
Fixed Temp. Operation, Automatic Start, Automatic Stop, Program Operation , Quick automatic stop
Additional Function
Total Operation Hours Timer (to 65535hrs.), calendar time ( 24 hours), calibration offset, Power discharge, CO2 discharge, power failure return mode, user configuration information
Heater Circuit Control
Triac Zero-cross Type
Safety Countermeasure
Self Diagnosis Function (Abnormal Temp. Sensor, Heater Wire Broken, Triac Short Circuit, Automatic Overheat Prevention), Independent Overheating Prevention, Key Lock, Electric Leakage Breaker
Internal dimensions(W×D×H mm)
450×490×450mm
External dimensions(W×D×H mm)
560×600×820mm
Power Source
AC115V11ASingle phase