System
Forced convection and ventilation
Operating temperature range
Room temp. +10 ~ 260℃
GB standard Temp. fluctuation
±0.1℃ (at 260℃)
Temp. adjustment accuracy
±0.1℃ (at 260℃)
Temp. distribution accuracy
±1.5℃ (at 260℃)
Max. temp. reaching time
Approx. 60 min (to 260℃)
Composition Interior material
Stainless steel plate
Exterior material
Cold rolled steel plate with chemical proofing coating
Insulating material
Glass fiber Rock wool
Heater
Stainless steel heating tubes with fins 3.0KW
Fan blade/motor
Axial fan blades, motor 20w × 1
Cable port
I.D. 30 mm, located on the back
Other additional mechanisms
Exhaust dampers (manual)
Controllers Temp. control method
3-stage PID
Temperature setting method
Digital setting through special function menu keys and up/down keys
Setting temp. display
Red 4-digit LED digital display
Timer
1 min. ~ 99 h 59 min. and 100~999 h 50 min (including timer waiting function)
Operation functions
Fixed temp. operation, auto start, auto stop, program operation
Program mode
Program operation 6 modes with a total of 90 segments (30 segments×1, 15 segments×2, 10 segments×3)
Additional functions
Deviation correction, key lock, power failure compensation
Heater circuit control
SSR drive
Sensors
K thermocouple (Temp. controller and overheating protector)
Safety device
Self-diagnostic circuit (temperature sensor anomaly, heater disconnection protection, automatic overheating preventer, SSR short-circuit), independent overheating preventer, overcurrent leakage protection, key lock functions
Specifications Internal dimensions (width × depth × height mm)
600×600×600
External dimensions (width x depth x height mm)
1200×780×1000
Accessories
Shelf Stainless steel wire mesh plate 3pcs ,Supports 6 pcs